Early diagnosis of keratoconus using corneal biomechanics and OCT derived technologies

Abstract

Background

Early detection of keratoconus is essential for maximizing the potential of cross-linking treatments designed to halt keratoconus progression, minimizing the risks of iatrogenic ectasia as well as reducing the need for corneal transplantation. This review focuses on the progress that has been made in the early detection of keratoconus using biomechanical and topographical properties derived from three different technologies, namely the ocular response analyser (ORA), corneal visualization Scheimpflug tonometer (Corvis ST) and optical coherence tomography (OCT).

Method

A PubMed search was performed using the keywords of ‘early keratoconus’, ‘subclinical keratoconus’, ‘forme fruste keratoconus’, ‘very asymmetric ectasia with normal topography/tomography’ and ‘ocular response analyser’ and/or ‘Corvis ST’/‘corneal visualized Scheimpflug tomographer/tomography’ and/or ‘optical coherence tomography/tomographer’.

Results

The integration of biomechanical parameters and corneal morphological data from the topography/tomography or OCT, or the assessment of bilateral asymmetry, has demonstrated improvement in the accuracy of diagnosing early-stage keratoconus.

Conclusions

As measurement principles differ depending on the technique used for keratoconus assessment, comprehensive metrics may be needed to reflect subtle anterior or posterior corneal changes and help identify eyes with very early ectasia. Although clinical experts have always, and will most likely, continue to play a pivotal role in decision-making for early keratoconus diagnosis, future developments in technology and AI may lead to enhanced early detection in the future.

Background

Keratoconus is a bilateral condition [1], characterized by corneal elevation, reduced corneal thickness, corneal biomechanical weakening, irregular astigmatism and progressive vision loss [2]. The associated risk factors include eye rubbing, a family history of keratoconus, allergy, asthma, and eczema [3]. The early detection of keratoconus can result in a better long-term prognosis, and it is crucial that keratoconus is excluded prior to refractive surgery [4]. Performing laser ablation surgery on these candidates could result in severe progressive iatrogenic ectasia [5, 6]. The advent of corneal topography and tomography has improved the early diagnostic ability of clinicians in identifying corneal ectasia, which may correspond to the higher keratoconus prevalence rates reported in recent years; up to 1.38 per 1000 population [3].

Nevertheless, it is still a challenge to differentiate borderline cases of keratoconus from normal corneas using tomography and topography, especially in small corneas where the 8 mm diameter best-fit sphere is incompatible with the size of the cornea and can produce a falsely elevated posterior surface [7]. This is especially poignant when considering different ethnic groups, as corneal diameter has been shown to be 0.2–0.5 mm narrower in Asians versus whites [7]. A further complication is that some patients with keratoconus may present typical clinical symptoms in one eye only, with the fellow eye showing no sign of abnormalities in corneal topography/tomography or slit-lamp examinations; this is referred to as forme fruste keratoconus (FFKC). In such cases, biomechanical abnormalities may be observed despite the appearance of normal tomography and topography [8]. According to the Global Consensus on Keratoconus and Ectatic Corneal Diseases [1], diagnosing early or subclinical keratoconus involves observing posterior corneal elevation abnormalities. This is supported by a number of studies that have highlighted the prime importance of posterior corneal changes [9], and the limited ability of detecting early keratoconus with simple Placido disc topography and keratometry which does not incorporate posterior corneal changes [10]. However, studies focusing on higher-order aberrations suggest that the primary contribution comes from the anterior corneal surface for detecting subclinical keratoconus [9, 11,12,13], and changes in the densitometry in the anterior central zone could be useful in detecting early-stage keratoconus as well [14, 15]. Therefore, whether the initial signs of keratoconus in the early stages can be identified at the anterior or the posterior corneal surface remains controversial.

Terminology has been problematic in this area. For example, the term ‘early-stage keratoconus’ may encompass FFKC or very asymmetric ectasia (VAE) [where one eye has clinical keratoconus and the fellow eye appears topographically normal (VAE-NT)], subclinical keratoconus (SKC), keratoconus suspect, pre-keratoconus, or mild keratoconus. In this review, we have used the most common definitions of SKC and FFKC. SKC is defined as ‘an eye with topographic signs of keratoconus and/or suspicious topographic findings under normal slit-lamp examination and keratoconus in the fellow eye, whilst FFKC is defined as ‘an eye with normal topography, normal slit-lamp examination, and keratoconus in the fellow eye’ [16, 17]. This definition would suggest that FFKC is less progressed than SKC towards KC. However, in a survey of 33 studies on SKC and 22 studies on FFKC, it was not possible to identify any specific examination that clearly differentiated SKC from FFKC [16]. For instance, keratoconus percentage (KISA%) values between 60% and 100% accounted for approximately 9% in both SKC and FFKC. On the other hand, all studies reported high incidences of keratoconus in the fellow eye, i.e., 72.72% in SKC and 77.27% in FFKC as a diagnostic criterion, indicating that a bilateral early diagnosis of the disease without clinical expression is a challenge [16]. In recent years, the term VAE-NT has been used synonymously with FFKC to describe patients with defined clinical ectasia in one eye and normal topography/tomography in the fellow eye [18].

This review focuses on the progress that has been made in the early detection of keratoconus (FFKC, VAE-NT, SKC) using biomechanical and topographical properties derived from three different technologies, namely the ocular response analyser (ORA), corneal visualization Scheimpflug tonometer (Corvis ST) and optical coherence tomography (OCT). A comprehensive search was performed in PubMed using the keywords of ‘early keratoconus’, ‘subclinical keratoconus’, ‘forme fruste keratoconus’, ‘very asymmetric ectasia with normal topography/tomography’ and ‘ocular response analyser’ and/or ‘Corvis ST’/‘corneal visualized Scheimpflug tomographer/tomography’ and/or ‘optical coherence tomography/tomographer’. Only those papers which clearly defined and characterised their populations of early keratoconus with slit-lamp findings, corneal topography and/or corneal tomography properties were included in the review.

Main text

Ocular response analyser (ORA)

The first in vivo measurements of corneal biomechanical response came with the introduction of the ORA (Reichert Ophthalmic Instruments, Buffalo, NY, USA) in 2005 [19]. The system was designed to improve the accuracy of intraocular pressure measurements by taking into account central corneal thickness (CCT) and the biomechanical properties of the tissue [19, 20]. The ORA is a noncontact tonometer which uses a collimated rapid air puff to indent the central 3–6 mm of the cornea and an advanced electro-optical system to monitor the bi-directional movement of the cornea in response to the air puff (Fig. 1). Using a reflected infrared signal, the system records the pressure as the cornea reaches its first applanation point during indentation, and then again, as it reaches its second applanation point whilst returning to its original shape [19, 21]. The difference between these two applanation pressures is thought to be caused by the viscoelasticity of the cornea and provides an estimated measure of corneal hysteresis (CH), which is an assessment of the cornea’s ability to absorb and dissipate energy or force. The ORA software also records a corneal resistance factor (CRF). Whilst CH principally relates to the viscous properties of the cornea, CRF is thought to be dominated by the elastic properties of the tissue [22].

Fig. 1
figure 1

Illustration of peaks measured with the ocular response analyser in a myopic eye. CRF, corneal resistance factor; CH, corneal hysteresis; p1area, upper 75% area of first peak; H1, height of first peak; H2, height of second peak; W1, width of peak 1 at point of the 25% of the base region W2, width of peak 2 at point of 25% of the base region

Schweitzer et al. demonstrated that the ORA could provide additional information to assist with the screening of FFKC. Compared with normal eyes, FFKC eyes showed significantly lower values of CH (9.1 ± 1.8 vs. 10.3 ± 1.9 mmHg) and CRF (9.2 ± 1.8 vs. 11.1 ± 2.0 mmHg). Further, the force and time needed to reach applanation was significantly lower in eyes with FFKC [23]. Whilst Ayar et al. reported similar results, showing differences between FFKC eyes and normal eyes in both CH (8.3 ± 1.6 vs. 9.8 ± 1.6 mmHg) and CRF (7.8 ± 1.2 vs. 9.9 ± 1.5 mmHg) [24], others have reported differences in the mean CRF between FFKC and normal eyes (7.8 ± 1.4 vs. 10.2 ± 1.7 mmHg) without any statistical difference in the mean CH [25, 26]. The reason for the difference between studies is most likely due to differences in the classification of the groups, as summarized in Table 1.

Table 1 Previous studies on early keratoconus with ocular response analyser (ORA)

Receiver operating characteristic (ROC) curves can be used to determine the predictive accuracy of the test parameters, as described by the area under the curve, and to calculate the sensitivity and specificity of these parameters. Using ROC analysis, Ayar et al. showed that area under the ROC curve (AUROC) values for CH and CRF needed to distinguish FFKC from normal eyes were 0.768 and 0.866, respectively [24]. However, Kirgiz et al. reported slightly higher AUROC values of CH (0.85) and CRF (0.90) to distinguish between the two groups [27]. In a study by Luz et al. investigating the use of different ORA exported signals, waveform parameters, and Pentacam HR metrics to distinguish FFKC from normal corneas, it was found that the Pentacam HR Belin/Ambrósio enhanced ectasia display (BAD), derived from corneal pachymetry, showed the highest predictive value (AUROC, 0.910 ± 0.057), followed by the ORA waveform measurement of the upper 50% area of the first peak (p1area1) (AUROC, 0.717 ± 0.065). However, further analysis revealed that FFKC screening could be significantly enhanced using a combined biomechanical and tomography approach (AUROC, 0.953 ± 0.024), rather than comparing individual parameters [28].

In a study of 61 patients with confirmed definite keratoconus in one eye and asymmetric contralateral corneal ectasia with normal topography (ACE-NT) in the other, it was shown that CRF (0.866) had a higher detection ability (cut-off ≤ 9.6) than CH (0.826) (cut-off ≤ 9.9) for differential diagnosis of normal from ACE-NT eyes [29]. Although most studies have found differences in CH and CRF between early keratoconus and normal eyes, a few studies have found no differences [30]. In VAE-NT eyes, the tomographically normal partner eye showed a significantly more pathological CH (8.5 ± 1.5 mmHg) and CRF (8.3 ± 1.5 mmHg) than the normal control group [31]. Atalay et al. showed that a combined biomechanical and tomographic approach using ORA CH and Pentacam Belin/Ambrósio enhanced ectasia display total deviation (BAD-D) index was optimal for detecting SKC, with the resulting AUROC, sensitivity and specificity (0.948, 87.1%, and 91.4% respectively) being higher than any other individual ORA or Pentacam parameter [32]. The multivariate analysis of ORA signals did not surpass simpler models in early keratoconus detection, and there was considerable overlap between normal and ectatic eyes, irrespective of the analysis model [30, 33].

As outlined above, multiple retrospective and prospective studies [31, 32, 34, 35] have shown that developing a combined biomechanical and tomographic model with multiple variables, may further improve detection of early keratoconus. With increasing complexity of diagnostic technologies, artificial intelligence (AI) has emerged as a promising tool to enhance keratoconus detection and improve accuracy, efficiency, and scalability [36, 37]. However, the development of AI models requires large and diverse data sets, meticulous planning of data inclusion criteria to avoid bias, and detailed database searching, data preprocessing, algorithm selection, testing and validation. A recent review of AI algorithms demonstrated the high degree of accuracy that AI can provide for the detection of manifest keratoconus (~ 98%) [38]. However, the accuracy for detecting subclinical keratoconus was found to be slightly lower (~ 90%), presenting a greater risk of missed diagnosis in patients without clinical signs. As AI systems continue to develop and improve in accuracy for the detection of early keratoconus, it was envisaged that they will become increasingly incorporated into clinical practice.

Corneal visualization Scheimpflug tonometer (Corvis ST)

The Corvis ST (Oculus, Wetzlar, Germany) is a noncontact tonometer that uses an ultra-high-speed Scheimpflug camera to capture over 4,300 frames per second, allowing the monitoring of corneal response to a metered, collimated air pulse with symmetrical fixed profile and a fixed maximal internal pump pressure of 25 kPa. The Scheimpflug camera has a blue light LED (475 nm, UV free) source and horizontally covers 8.5 mm of a single slit. Recording measurement time is 30 ms [42]. This also allows a more detailed description of the dynamic deformation process in comparison with the ORA system [43].

Corneal biomechanical properties using the Corvis ST can be presented as dynamic corneal response parameters associated with biomechanically adjusted intraocular pressure (bIOP), including applanation length (A1L and A2L), applanation velocity (A1V and A2V) and applanation time (A1T and A2T) at the first and second applanation, the distance between the two peaks, and axial displacement of the apex of the cornea [deformation amplitude (DA) and central radius of the cornea at the highest concavity phase, stiffness parameter at the first applanation (SP-A1), integrated radius (IR), the Ambrósio relational thickness (ART), and deformation amplitude ratio (DA ratio 2 = DA at the apex/average of DA at 2 mm around the center in the horizontal directions]. In addition, the combined biomechanical parameters include the corneal biomechanical index (CBI), tomographic and biomechanical index (TBI), and most recently, the stress-strain index (SSI).

ART is the division between corneal thickness at the thinnest point and the Pachymetric Progression Index. A lower value indicates a thinner cornea and/or a faster thickness increase toward the periphery [44]. SP-A1 uses the displacement between the undeformed apex and the position of first applanation. SP-A1 is defined as the resultant pressure (Pr) divided by deflection amplitude at A1. Resultant pressure (Pr) is defined as the adjusted pressure at A1 (adj AP1) minus a biomechanically adjusted IOP value (bIOP), resulting in the following equation: SP-A1 = (adjusted AP1–bIOP)/A1 deflection amplitude. The spatial and temporal profiles of the Corvis ST air pulse are measured using hot wire anemometry [44, 45]. CBI is a logistic regression algorithm that combines different biomechanical parameters to optimize the separation keratoconus eyes and normal eyes. It includes several dynamic corneal response parameters: deformation amplitude ratio (DA ratio) at 1 and 2 mm, applanation 1 velocity, the standard deviation of deformation amplitude at highest concavity, Ambrósio relational thickness to the horizontal profile (ARTh), SP-A1, and IR. With a cutoff value of 0.5 using CBI, 98.2% of the cases were correctly classified with 100.0% specificity and 94.1% sensitivity in differentiating definite keratoconic patients from the normal population [44]. TBI is calculated using an AI approach to optimize ectasia detection. By combining tomographic data from the Pentacam HR with biomechanical data from the Corvis ST, one can further improve sensitivity and specificity in the detection of patients with a significant risk for developing ectasia after refractive surgery [35]. The SSI was generated based on predictions of corneal behaviour using numerical finite element modelling, which simulates the effects of IOP and the Corvis ST air puff to estimate the stiffness of the material [46]. The newly developed SSI II (SSI map) provides an estimation of the regional variation of biomechanical stiffness across the corneal surface and it is anticipated that these maps could be particularly useful in understanding keratoconus development and progression [47].

Corvis ST in FFKC

Although some clinicians may misuse the term of ‘topography’ and ‘tomography’, the most widely used pre-surgery screening technique of detecting corneal topographic/tomographic information is relying on both the anterior and posterior surface, by Scheimpflug based and Placido-disk based instruments. Hence, in FFKC, using these modalities, the topographic/tomographic indices look the same as normal eyes, and the only distinctive parameter between them is the abnormal ectatic contralateral eye [29]. Therefore, the biomechanical parameters might be helpful in such cases.

Hwang et al. [48] reported that the maximum ART (ARTmax) yielded an AUROC of 0.739 (sensitivity of 56.7% and specificity of 88.3%), to distinguish the less affected eye of VAE (the same definition as VAE-NT) from normal eyes. Awad et al. [49] also found that ART (AUROC: 0.88) was a highly sensitive objective parameter in FFKC cases. However, Shajari et al. [50] demonstrated ARTmax (AUROC: 0.613) performed less well in differentiating populations at early stages of keratoconus. They proposed that using index of height decentration and index of vertical asymmetry were better markers in the early stages of keratoconus, and when the disease progresses, the BAD-D index is better suited to diagnose ectasia [50].

Some researchers have found that corneal biomechanical properties are highly sensitive for the detection of early keratoconus even in the absence of topographic abnormalities. After controlling for CCT and bIOP, the values of corneal deflection amplitude during the first applanation, A1L, corneal deflection amplitude during the second applanation, and maximum deformation amplitude increased in the FFKC compared with the normally thin cornea [51]. Tian et al. also found that A1T was faster, and the SP-A1 and CBI were significantly softer in FFKC cases compared to the corneal thickness and IOP in age-matched healthy corneas [52]. Among the top five indices of the AUROCs for detecting early keratoconus (similarly to FFKC), the corneal biomechanical-related index accounted for 80%, including the A1 dArc length (AUROC: 0.901), highest concavity radius (AUROC: 0.879), A2T (AUROC: 0.877), and TBI (AUROC: 0.874) [53]. The BAD-D also provided a high predictive value (AUROC: 0.91 ± 0.057) [28]. Another study in a Chinese population confirmed the discriminatory values of TBI (AUROC: 0.928, cutoff: 0.38, Youden index: 0.753), and CBI (AUROC: 0.860, cutoff: 0.27, Youden index: 0.642) for distinguishing FFKC from normal eyes [54]. Significant differences were found for A2L, A1V, A2V, and TBI between subclinical keratoconus (similarly to FFKC) and normal eyes. TBI showed the highest AUROC (0.790; cutoff: 0.29; sensitivity: 67%; specificity: 86%) in distinguishing subclinical keratoconus from normal eyes [55].

Although corneal biomechanical stability was found to be significantly lower in FFKC eyes than in normal eyes, with significantly decreased SP-A1 and increased TBI, the AUROCs of SP-A1 and TBI for identifying FFKC were lower than 0.7 [56]. Wang et al. found that comparable AUROC and partial AUROC was observed between the CBI (AUROC: 0.785; pAUROC: 0.079) and BAD-D (AUROC: 0.757; pAUROC: 0.068) for detecting FFKC with sensitivities of 63.2% and 52.6%, given a common specificity of 80.3%, which is lower than in the advanced keratoconus cases [57]. Tian et al. established the keratoconus diagnosis model using backpropagation neural network, and found that the predicted value (AUROC: 0.877) was more sensitive in the detection of FFKC than CBI (AUROC: 0.610) and TBI (AUROC: 0.659) [52]. Luz et al. [28] also found that the logistic regression model yielded the highest accuracy (AUROC: 0.953, sensitivity: 85.71%, specificity: 98.68%). This model incorporated variables such as BAD-D, ARTmax, and the thinnest point-related elevation on both the anterior and posterior corneal surfaces [28].

In a study involving 137 patients, SSI was found to differ significantly between keratoconus, FFKC and normal eyes, indicating an independent decrease in corneal stiffness in eyes with keratoconus. However, the AUROC and Youden index of the SSI were not as good as TBI (AUROC: 0.928), BAD-D (AUROC: 0.926) and CBI (AUROC: 0.860) in detecting FFKC [54]. After correcting for CCT and bIOP, SSI II and ART were significantly higher, and CBI was significantly lower in the normal group than in the FFKC group, SKC group and clinical keratoconus groups. AUROC of SSI II was significantly higher than all other Corvis parameters in distinguishing normal eyes from FFKC, followed by ART and CBI [58].

In a secondary statistical analysis of 80 publications utilizing Pentacam or Corvis ST parameters, it was found that except for CBI, SPA1 was the only Corvis ST output parameter sensitive to FFKC (AUROC: 0.87, sensitivity: 0.71, and specificity: 0.85). SP-A1 was not inferior to the CBI and may be the earliest Corvis ST output to reflect changes in corneal biomechanics during keratoconus progression. Furthermore, it was found that most of the 20 sensitive diagnostic parameters screened that were related to the thinnest point of the cornea, such as thickness progression parameters, height parameters based on the thinnest point, and even thinnest point of the cornea itself, were selected as the sensitive diagnostic parameters of FFKC [59].

Covis ST in subclinical keratoconus

Although clinically, both SKC and FFKC are not evident on slit-lamp examination, the biomechanical parameters A1T, IR, and TBI are significantly varied between the SKC and FFKC, suggesting that these parameters are more powerful for detecting subtle changes in corneal biomechanical properties than SP-A1. Overall, the biomechanical properties of SKC are weaker than FFKC [60].

In an age controlled normal vs. SKC study, Peris-Martinez et al. evidenced the usefulness of the biomechanical parameters provided by Corvis ST in detecting subclinical keratoconus, showing statistically significant differences in maximum deformation amplitude, highest concavity radius, A2L and A2V [61]. In addition, others found that A1L, A2L, radius of the inward-bended cornea, and deflection length at the highest concavity parameters demonstrated statistically significant differences [62]. Table 2 provides a summary of studies examining early keratoconus detection with Corvis ST.

Table 2 Previous studies on early keratoconus with Corvis ST

Random forest machine learning techniques could advance the classification and prediction of SKC through developing an algorithm for mining the metrics generated by the Pentacam HR and Corvis ST. The random forest approach is deemed a good model for classifying SKC, demonstrating a specificity of 93% and sensitivity of 86% [63]. SP-A1 can become a critical determinant in classifying and identifying SKC when compared to maximum deformation amplitude radius at 2 mm and 1 mm, IR, ARTh and CBI, followed by A2T [63]. In line with this, SP-A1 and CBI were significantly different between normal control and SKC eyes, and the parameter with the highest diagnostic efficiency was SP-A1 (Youden index: 0.40, AUROC: 0.753), followed by CBI (Youden index: 0.38, AUROC: 0.703), and IR (Youden index: 0.33, AUROC: 0.668) [64]. Heidari et al. [65] reported close AUROCs for SP-A1 (0.779), and lower AUROCs for ART (0.718), CBI (0.758), and TBI (0.828), which were all inferior to the Sirius symmetry index of back (0.908) and Pentacam HR I-S value (0.862) in differentiating SKC from normal eyes.

A significant difference was also found in the BAD-D and TBI between normal and SKC eyes [66]. However, whilst some found TBI to be optimal for the detection of SKC (AUROC: 0.925 for TBI vs. 0.786 for BAD-D) [66], others found that BAD-D provided a slightly better diagnostic performance (AUROC: 0.944 for TBI vs. 0.965 for BAD-D) [67]. Interestingly, AI-driven approaches, when integrated with advanced tools like the Scheimpflug rotation camera-based elevation map, Placido-disk-based keratometry, and extensive databases, show great promise in identifying early-stage keratoconus. However, the role of clinical experts remains irreplaceable, especially when making critical decisions prior to refractive surgery. That said, there may still be some concerns regarding the level of experience among clinical specialists in leveraging these technologies effectively [68].

Corvis ST in VAE-NT eyes

Objective criteria for considering normal topography was rigorously applied for defining cases of VAE-NT, and included objective front surface curvature metrics derived from the Pentacam HR, such as a keratoconus percentage index (KISA%) score lower than 60% and a paracentral inferior-superior (I-S value) asymmetry value at 6 mm (3-mm radii) less than 1.45 D [35, 39].

In a recent study, in which 14 patients with VAE-NT were enrolled (although the tomographically normal fellow eyes of keratoconus patients are rare), regular CBI values (0–0.249) were found in 6/14 of the VAE-NT eyes examined. The mean TBI was 0.47 ± 0.22 (range: 0.22–0.84) with regular TBI values (0–0.249) recorded in only 2/14 patients. The sensitivity of both, CBI (99.1%) and TBI (99.6%) in detecting tomographic abnormal keratoconus can be considered very high [69]. Furthermore, another study comparing the use of biomechanical indices generated by the ORA and Corvis ST (a dynamic Scheimpflug analyser) to distinguish between normal eyes and those with very asymmetric ectasia [defined as unilateral keratoconus, where one eye has clinical keratoconus and the fellow eye appears topographically normal (VAE-NT) or topographically and tomographically normal (VAE-NTT)], it was found that the Corvis biomechanical index had the highest AUROC (0.979), followed by ORA CRF (0.865) and CH (0.824) [70]. The CBI (cutoff 0.2) showed a sensitivity of 100.0% and 70.9%, respectively, and a specificity of 93.1% for distinguishing normal eyes from VAE-NT and VAE-NTT, whereas the values for ORA CRF and CH were much lower [70].

TBI combines Scheimpflug-based corneal tomography and biomechanics for enhancing ectasia detection. A retrospective analysis of Corvis HR data from the eyes of 684 patients revealed the AUROCs for TBI, BAD-D, and CBI to be 0.985, 0.839, and 0.822 in discriminating VAE-NT eyes from normal controls [35]. This superior performance of TBI over CBI and BAD-D in distinguishing VAE-NT from normal corneas is also supported by several groups [71,72,73]. Although high accuracy (> 85%) has been reported in many studies, a number of cutoff values exist ranging from 0.16 [66], 0.24 [29], 0.29 [35], 0.63 [71], 0.72 [34, 72], 0.295 [74], to 0.259 [75].

Fraenkel et al. examined ORA and Corvis ST parameters in normal and VAE-NT eyes and provided evidence of pathological CH (8.5 ± 1.5 mm Hg) and CRF (8.3 ± 1.5 mm Hg) in the VAE-NT eyes. Although the average TBI value for the VAE-NT eyes (0.19 ± 0.25) did not differ significantly from the normal eyes, 5/26 VAE-NT eyes (19.2%) had a TBI of more than 0.29 and were considered pathological [31]. In larger study by Sedaghat et al. it was shown that CRF (AUROC: 0.866) had a higher detection ability than CH (AUROC: 0.826) in distinguishing normal eyes from VAE-NT eyes, and the discriminative analysis showed that the highest accuracy of the Corvis ST was related to TBI (0.966) [29].

However, controversial results were observed for differentiating normal and VAE-NT eyes. The AUROC for the BAD-D, CBI, and TBI were 0.668, 0.660, and 0.751, respectively. The TBI cut-off of 0.259 provided a sensitivity of 52.17% and a specificity of 88.57% [75]. Nine VAE-NT cases (39.1%) were found to exhibit normal values for BAD-D, CBI, and TBI. Moreover, 40% of VAE-NT eyes were classified as normal through the use of BAD-D, CBI, and TBI [75]. Although some of these cases may truly represent unilateral ectasia, further advances are needed to enhance ectasia detection and characterize the susceptibility for ectasia progression. In another observational study, of more than 900 participants, 34 eyes with VAE-NT (7.4%) were identified and included in the analysis. Biomechanical analysis demonstrated a mean CBI of 0.28 ± 0.26 and a mean TBI of 0.34 ± 0.30. Out of the 34 eyes, 16 (47%) normal CBI values, 13 (38%) regular TBI and 7 (21%) regular TBI and CBI were observed. The sensitivity of CBI and TBI in detecting a tomographically normal keratoconus fellow eye was 53% and 62%, respectively. TBI showed slightly higher sensitivity than the CBI (62% vs. 53%) for detecting keratoconus in a tomographically unremarkable keratoconus partner eye. Further, 21% of the keratoconus partner eyes could not be recognized as conspicuous, either by CBI or TBI [76]. Even though the SSI value decreased with keratoconus progression, no differences in SSI were observed between healthy individuals and VAE-NT cases, which may be due to the focal nature of keratoconus [77].

The index of height asymmetry and height decentration differed significantly between FFKC and mild keratoconus eyes and thin normal corneas. The index of height decentration also had sufficient strength (AUROC > 0.80) to differentiate FFKC and mild keratoconus from thin normal corneas eyes. The deflection amplitude of the first applanation showed a good potential to differentiate (AUROC > 0.70) FFKC and mild keratoconus from thin normal corneas [78]. Central astigmatism from the anterior corneal surface had an AUROC of 0.862 for distinguishing FFKC from normal eyes. This was associated with a cutoff point of 4.65, providing 73.5% sensitivity and 99.3% specificity. Conversely, other combined parameters, including CBI, TBI, and BAD-D, had low AUROCs to differentiating FFKC from normal eyes [79].

Zhang et al. [80] applied a stepwise logistic regression model to differentiate FFKC from normal eyes. The model was derived from the parameter ‘elevation of front surface in thinnest location’ from the Pentacam and deflection amplitude of the highest concavity and SP-A1 from the Corvis ST. The predictive accuracy of this model presented an excellent AUROC (0.965) with 100.0% sensitivity and 84.0% specificity, followed by TBI (0.885), elevation of front surface in thinnest location (0.874) and BAD-D (0.839) alone. However, the CRF and CH output by ORA did not improve the combined diagnosis, despite the corneal combination of morphological and biomechanical properties that optimized the diagnosis of FFKC [80].

In summary, the diverse screening methods used across various studies mean that participants in each study may reflect different stages of ectatic disease. The Corvis ST-derived parameters—whether basic, advanced, or integrated—exhibited varying performance, leading to a wide spectrum of cutoff values for each index.

Development based on the current Corvis ST indices

A notable limitation of the Corvis ST in the screening of keratoconus is that it automatically images a single, central 8.5 mm horizontal section of the cornea [55, 81]. Given that a common location of focal thinning in keratoconus is the inferotemporal site, the Corvis ST may therefore fail to capture the focal area of corneal biomechanical compromise in keratoconus [82, 83]. Kenia et al. found that in a subclinical keratoconus group, combining epithelial mapping with corneal biomechanical parameters could help improve the efficacy in the diagnosis of SKC [84]. Furthermore, a retrospective comparative study of 33 clinically unaffected eyes characterised by normal corneal topography and biomechanical properties, revealed abnormal pachymetric patterns using Pentacam within the central cornea, alongside subtle morphological alterations evident in early-stage keratoconus with preserved biomechanics [85]. These findings collectively underscore the potential of morphological parameters, specifically those related to epithelial characteristics and pachymetric distribution, to furnish supplementary diagnostic insights into the early stages of keratoconus.

Additionally, due to the distribution range of many biomechanical parameters overlapping between normal individuals and SKC patients, it was hypothesised that introducing biomechanical interocular asymmetry assessment may reduce the false-negative rate and improve the sensitivity of diagnosing SKC [86]. Indeed, this was demonstrated using the binocular assisted biomechanical index, which produced an AUROC of 0.998 (sensitivity: 97.8%, specificity: 99.2%; cutoff: 0.401), which was statistically higher than CBI (AUROC: 0.935, sensitivity: 85.6%) [86].

Finally, significant differences in the values of SP-A1 and SSI and ART were noticed between the Caucasians and the Chinese [87]. A relatively high false rate for keratoconus suspects was also found in the Chinese population [54]. The new CBI optimized for Chinese patients (cCBI) was investigated and showed an AUROC of 0.990. With a cutoff value of 0.5, it produced 93.7% specificity and 95.9% sensitivity in the training dataset for distinguishing healthy eyes from typical keratoconus eyes. Conversely, the original CBI produced an AUROC of 0.981 with 75.4% specificity and 97.9% sensitivity [88]. It is important to note that in this new cCBI study, they excluded unilateral keratoconus, FFKC, and SKC from the databases when creating and validating the cCBI. More studies are warranted to test the capability of cCBI to separate healthy individuals from those with FFKC and other early stages of keratoconus in Chinese patients.

Optical coherence tomography (OCT)

Anterior segment OCT is another noncontact high-resolution imaging technique available for three-dimensional anterior segment imaging in keratoconus [90]. Since its introduction in the 1990’s, significant advances have been made in OCT imaging of the cornea. Time-domain OCT was initially developed for capturing cross-sectional images of the eye’s anterior segment [91]. Subsequently, 840 nm spectral-domain OCT was introduced, with a faster image acquisition and better resolution; however, this advancement led to more limited image depth ranges and horizontal scan widths. Additionally, other variations of spectral-domain OCT with improved axial resolutions, termed ultra-high-resolution OCT have been developed. Currently, a 1310 nm swept-source OCT based on the Fourier-domain OCT type allows for more precise and quicker reconstruction of three-dimensional anterior segment images [90]. A summary of currently commercially available OCT devices is shown in Table 3.

Table 3 Summary of currently commercially available anterior segment optical coherence tomography (OCT)

OCT in the early diagnosis with corneal morphology analysis

At present, OCT excels in the identification and digitalization of both corneal surfaces with good repeatability and high reliability [94], providing curvature and elevation maps of the anterior and posterior cornea and corneal thickness (along with corneal pachymetry maps) [95]. Additionally, the advantages of OCT lie in providing clearer images of corneas with opacities using longer wavelength scanning beams [96,97,98]. Further, the Fourier analysis of corneal power with OCT can quantitatively evaluate corneal irregular astigmatism in the various stages of keratoconus, including advanced keratoconus cases with corneal opacities [99]. Therefore, corneal analysis using OCT can quantify the refractive component, corneal thickness, and irregular astigmatism of the cornea across different stages (very mild to advanced) and may also be used for longitudinal observation.

Parameters for early detection in FFKC, have been studied for both anterior and posterior corneal surfaces with OCT. In previous studies, variables used to define FFKC were reported as keratoconus in the fellow eye in 77.27%, normal topography in 59.09%, and normal slit-lamp examination in 40.90%. Among these studies, 90.90% used more than one parameter to define FFKC. However, topographic signs and parameters used to define FFKC differ between groups, making it difficult to unify them, and will be noted in Table 4 [16].

Table 4 Previous studies on forme fruste keratoconus (FFKC) with optical coherence tomography (OCT)

For RTVue (a Fourier-domain system from Optovue Inc., Fremont, CA, USA), other than epithelial thickness, ectasia index, computed from pachymetry and posterior surface mean curvature, has been studied by Pavlatos et al. in FFKC [100]. The results have shown that the subtle early signs of keratoconus may vary between cases, and combining corneal thickness and posterior curvature information is more effective than relying on only one of these measurements, especially in FFKC. Hwang et al. [48] have reported a better resolution in combining anterior curvature and asymmetry indices from Scheimpflug imaging [Pentacam HR (Oculus, Wetzlar, Germany)] with regional total thickness and epithelial thickness variability metrics from OCT (RTVue-100) for FFKC. Combining five Scheimpflug metrics provided better results, while combining 11 OCT thickness metrics yielded the best results from OCT. Moreover, combining 13 total Scheimpflug and OCT metrics gave the highest results. The most impactful variables included epithelial thickness variability and total focal corneal thickness variability from OCT, along with anterior curvature and topometric indices from Scheimpflug technology. Interestingly, no posterior corneal indices were significant in this model.

For the ANTERION (a swept-source system from Heidelberg Engineering GmbH, Heidelberg, Germany), curvature parameters of the anterior and posterior corneal surface have been studied. Saad et al. examined anterior and posterior curvature and thickness parameters. They reported an anterior curvature–derived parameter, the inferior-superior asymmetry value (I-S value) [39], yielded the highest AUROC in discriminating FFKC from normal eyes, followed by the magnitude of inferior decentration of posterior steepest keratometry [95].

For the CASIA SS-1000/CASIA2 (a swept-source system from Tomey Corporation, Aichi, Japan), posterior corneal elevation has also been studied in FFKC. According to Fukuda et al., posterior corneal elevation was determined to demonstrate the most reliable predictive accuracy in detecting FFKC [101]. Kitazawa et al. and Itoi et al. calculated and demonstrated that the ratio of the anterior corneal surface area to the posterior corneal surface area, derived from elevation maps obtained through OCT, was useful for discriminating FFKC cases from normal eyes, indicating that an imbalance in the anterior and posterior corneal surface areas may reflect early-stage keratoconus [102, 103]. Regarding quantitative Fourier harmonic analysis of corneal power distribution, Maeno et al. have reported that posterior irregular corneal astigmatism exhibits the highest discrimination ability among anterior or posterior components for FFKC [99]. Automated single parameters and newly computed indices using the CASIA have been combined to further improve this accuracy [104]. Shiga et al. have studied Fourier parameters with the CASIA as well, indicating the combination of the posterior corneal asymmetric component and central corneal thickness in FFKC discrimination [105]. Moreover, when the posterior corneal asymmetric component from anterior segment optical coherence tomography (AS-OCT) and the parameter SP-A1 from corneal biomechanical assessment (Corvis ST, Oculus Optikgeräte) were combined, it provided higher AUROC, although no significant differences in AUROC across diagnoses when comparing the use of each individual device versus their combined uses (integrated parameters versus AS-OCT, P = 0.081; integrated parameters versus Scheimpflug-based tonometer, P = 0.234).

Diagnosing FFKC by integrating biomechanical properties with morphology analysis has shown superiority compared to diagnosis based on a single device. Combining parameters obtained from multimodal imaging, such as Scheimpflug and OCT epithelial thickness mapping [48, 106], or tomographic and biomechanical assessments [35, 80] result in better detection compared to individual parameters. This indicates that corneal biomechanics can be a complementary tool for FFKC identification in addition to tomographic parameters. In fact, rather than a single individual Fourier posterior asymmetry component, enhanced subclinical ectasia detection capability by combining posterior Fourier indices and biomechanical properties might be possible.

With approval from the Institutional Review Board/Ethics Committee of Osaka University Hospital (registration number: 09297–20), we re-examined our previously published AS-OCT from 50 FFKC eyes of 50 patients and 44 controls [99] alongside Scheimpflug-based tomography and biomechanical data obtained from the same patients using Pentacam HR and Corvis ST, respectively [99]. After determining BAD-D using Scheimpflug-based tomography (Pentacam HR), CBI and TBI using corneal biomechanical assessment (Corvis ST), and the posterior asymmetry component from the Fourier analysis with OCT (CASIA2), we compared the AUROC, sensitivity, and specificity for these four indices and variable combinations. Despite this, no individual parameter from Scheimpflug, OCT, or biomechanical assessment yielded an AUROC higher than 0.80 (Fig. 2). Fourier posterior asymmetry with OCT gave the highest AUROC (0.778) [99] (see Table 5). When parameters were combined using multiple regression analysis, the combination of posterior corneal asymmetry and CBI significantly improved the discrimination accuracy for FFKC (Fig. 2) (P = 0.024 between ROCs for the Fourier posterior asymmetry component and the combined metric). This yielded the highest AUROC (0.863) almost equivalent to other combined parameters by Shiga et al., and improved sensitivity compared to single parameter detection (Table 5). Further studies combining both devices may be conducted, as differences in sample size could be contributing to the variation in these results.

Fig. 2
figure 2

Receiver operating characteristic (ROC) curves of single and combined parameters for forme fruste keratoconus (FFKC) discriminating ability. The discriminating ability accuracies of the Fourier indices were assessed between FFKC and control. ROC curves for each Fourier index for the central 6 mm of the cornea, the asymmetry component of the anterior and the posterior cornea, CBI, TBI, BAD-D, and combined posterior asymmetry and CBI (combined metric) are shown. FFKC eyes had the highest area under the ROC curve (AUROC) values for the combined metric. CBI corneal biomechanical index; TBI, tomographic and biomechanical index; BAD-D, Belin/Ambrósio enhanced ectasia display total deviation

Table 5 Optimum cut-off values with the highest sensitivity and specificity

OCT in the early diagnosis with corneal epithelial thickness mapping

Changes in total corneal thickness associated with keratoconus arise from alterations in both stromal and epithelial layers [107, 108]. In keratoconus, the epithelium exhibits thinning in the area overlying the cone, and this phenomenon has been confirmed through corneal epithelial mapping with OCT [109, 110]. This epithelial remodelling occurs in response to underlying stromal irregularities, with progressive stromal thinning contributing to the advancement of the disease. Recent progress have enabled thickness mapping of both epithelial and stromal thickness profiles [111,112,113].

Using RTVue, epithelial thickness mapping reportedly detect the initiating event of keratoconus that cannot be identified using topographic and tomographic analyses [109, 111, 114, 115]. Furthermore, more devices such as CASIA2 have recently been shown to effectively measure corneal layer thickness, providing valuable insights for detecting keratoconus [116]. Instruments such as ANTERION, MS-39, and Avanti also enable epithelial thickness measurements in eyes affected by keratoconus [110, 117].

Distinguishing FFKC from normal corneas warrants further study as a single corneal epithelial thickness parameter has proven inadequate for detecting the very early stages of keratoconus. To enhance diagnostic precision, efforts have been made to incorporate parameters from posterior cornea, corneal biomechanics or the standard deviation of corneal epithelial pattern [106, 114, 118]. Researchers have emphasized that combining multiple diagnostic modalities is critical for improving early detection. Additionally, custom-designed polarization-sensitive OCT based on swept-source OCT technology, has shown potential in visualizing the pathological changes in keratoconus including changes in the Bowman layer [119], and could be a valuable tool for identifying FFKC.

Conclusions

As highlighted in this review, the criteria for FFKC, SKC and VAE-NT often differ between studies and there is a lack of consistency regarding the terminology of early keratoconus detection. As measurement principles differ depending on the technique used for keratoconus assessment, comprehensive metrics may be needed to reflect subtle anterior or posterior corneal changes and help identify eyes with very early ectasia.

In this review, we primarily utilized AUROC values as a standard measure of diagnostic accuracy to demonstrate the efficacy of various diagnostic tools. However, it should be noted that incorporating additional statistical measures, such as positive predictive value and negative predictive value, may provide a more comprehensive evaluation of the clinical utility of these diagnostic tools.

In the foreseeable future, techniques such as OCT, Scheimpflug, non-contact tonometry (Corvis ST) and genetic analysis will undoubtedly continue to play a key role in the early detection of keratoconus. The advancement of these technologies and the further development of other less well investigated techniques such as Brillouin light-scattering microscopy, ultra-high-resolution ultrasound, optical coherence elastography and atomic force microscopy may lead to the earlier detection of keratoconus. Although it is crucial to acknowledge that clinical experts have always, and will most likely, continue to play a pivotal role in decision-making for early keratoconus diagnosis, it is envisaged that data interpretation will likely be enhanced by AI.

Availability of data and materials

Not applicable.

Abbreviations

A1L:

Length at the first applanation

A2L:

Length at the second applanation

A1T:

Time to reach the first applanation

A2T:

Time to reach the second applanation

A1V:

Velocity at the first applanation

A2V:

Velocity at the second applanation

ACE-NT:

Asymmetric contralateral corneal ectasia with normal topography

AI:

Artificial intelligence

AKC:

Asymmetric keratoconus

ART:

Ambrósio relational thickness

ARTh:

Ambrósio relational thickness to the horizontal profile

ARTmax:

Maximum Ambrósio relational thickness

AUROC:

Area under the receiver operating characteristic curve

BAD:

Belin/Ambrósio enhanced ectasia display

BAD-D:

Belin/Ambrósio enhanced ectasia display total deviation

bIOP:

Biomechanically adjusted intraocular pressure

CBI:

Corneal biomechanical index

cCBI:

CBI optimized for Chinese patients

CCT:

Central corneal thickness

CDVA:

Corrected distance visual acuity

CH:

Corneal hysteresis

Corvis ST:

Corneal visualization Scheimpflug tonometer

CRF:

Corneal resistance factor

DA ratio:

Deformation amplitude ratio

DA:

Deformation amplitude

FFKC:

Forme fruste keratoconus

H1:

Height from lowest to highest point in peak 1

H2:

Height from lowest to highest point in peak 2

IOP:

Intraocular pressure

IR:

Integrated radius

I-S value:

Inferior–superior asymmetry value

KC:

Keratoconus

KISA%:

Keratoconus percentage

Kmax:

Maximum keratometry

KSS:

Keratoconus Severity Score

OCT:

Optical coherence tomography

ORA:

Ocular response analyser

p1 area:

Upper 75% area of peak 1

Pr:

Resultant pressure

ROC:

Receiver operating characteristic curve

SKC:

Subclinical keratoconus

SP-A1:

Stiffness parameter at the first applanation

SSI:

Stress-strain index

TBI:

Tomographic and biomechanical index

VAE:

Very asymmetric ectasia

VAE-NT:

Very asymmetric ectasia where one eye has clinical keratoconus and the fellow eye appears topographically normal

VAE-NTT:

Very asymmetric ectasia where one eye has clinical keratoconus and the fellow eye appears topographically and tomographically normal

References

  1. Gomes JA, Rapuano CJ, Belin MW, Ambrósio R Jr, Group of panelists for the global Delphi panel of keratoconus and ectatic diseases. Global consensus on keratoconus diagnosis. Cornea. 2015;34(12):e38–9.

    Article  PubMed  Google Scholar 

  2. Zhao F, Du F, Zhang J, Xu J. Trends in research related to keratoconus from 2009 to 2018: a bibliometric and knowledge mapping analysis. Cornea. 2019;38(7):847–54.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Hashemi H, Heydarian S, Hooshmand E, Saatchi M, Yekta A, Aghamirsalim M, et al. The prevalence and risk factors for keratoconus: a systematic review and meta-analysis. Cornea. 2020;39(2):263–70.

    Article  PubMed  Google Scholar 

  4. Peyman A, Sepahvand F, Pourazizi M, Noorshargh P, Forouhari A. Corneal biomechanics in normal and subclinical keratoconus eyes. BMC Ophthalmol. 2023;23(1):459.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Mattila JS, Holopainen JM. Bilateral ectasia after femtosecond laser-assisted small incision lenticule extraction (SMILE). J Refract Surg. 2016;32(7):497–500.

    Article  PubMed  Google Scholar 

  6. Qin V, Saad A, Gatinel D. Unilateral post-LASIK ectasia and contralateral spontaneous ectasia. J Refract Surg. 2016;32(2):127–30.

    Article  PubMed  Google Scholar 

  7. Boyd BM, Bai J, Borgstrom M, Belin MW. Comparison of Chinese and North American tomographic parameters and the implications for refractive surgery screening. Asia Pac J Ophthalmol (Phila). 2020;9(2):117–25.

    Article  PubMed  Google Scholar 

  8. Vinciguerra R, Ambrósio R Jr, Roberts CJ, Azzolini C, Vinciguerra P. Biomechanical characterization of subclinical keratoconus without topographic or tomographic abnormalities. J Refract Surg. 2017;33(6):399–407.

    Article  PubMed  Google Scholar 

  9. Santodomingo-Rubido J, Carracedo G, Suzaki A, Villa-Collar C, Vincent SJ, Wolffsohn JS. Keratoconus: an updated review. Cont Lens Anterior Eye. 2022;45(3):101559.

    Article  PubMed  Google Scholar 

  10. de Sanctis U, Loiacono C, Richiardi L, Turco D, Mutani B, Grignolo FM. Sensitivity and specificity of posterior corneal elevation measured by Pentacam in discriminating keratoconus/subclinical keratoconus. Ophthalmology. 2008;115(9):1534–9.

    Article  PubMed  Google Scholar 

  11. Koh S, Inoue R, Maeno S, Mihashi T, Maeda N, Jhanji V, et al. Characteristics of higher-order aberrations in different stages of keratoconus. Eye Contact Lens. 2022;48(6):256–60.

    Article  PubMed  Google Scholar 

  12. Koh S, Matsuo R, Inoue R, Miyazato A, Asonuma S, Maeno S, et al. A comprehensive wavefront assessment of keratoconus using an integrated Scheimpflug corneal tomographer/Hartmann-Shack wavefront aberrometer. Eye Contact Lens. 2024;50(1):16–22.

    PubMed  Google Scholar 

  13. Heidari Z, Mohammadpour M, Hashemi H, Jafarzadehpur E, Moghaddasi A, Yaseri M, et al. Early diagnosis of subclinical keratoconus by wavefront parameters using Scheimpflug, Placido and Hartmann-Shack based devices. Int Ophthalmol. 2020;40(7):1659–71.

    Article  PubMed  Google Scholar 

  14. Koc M, Tekin K, Tekin MI, Uzel MM, Kosekahya P, Ozulken K, et al. An early finding of keratoconus: increase in corneal densitometry. Cornea. 2018;37(5):580–6.

    Article  PubMed  Google Scholar 

  15. Ozkan HH, Koc M, Kiziltoprak H, Tekin K, Aydemir E. Evaluation of topographic, tomographic, topometric, densitometric, and aberrometric features of cornea with Pentacam HR system in subclinical keratoconus. Int Ophthalmol. 2021;41(5):1729–41.

    Article  PubMed  Google Scholar 

  16. Henriquez MA, Hadid M, Izquierdo L Jr. A systematic review of subclinical keratoconus and forme fruste keratoconus. J Refract Surg. 2020;36(4):270–9.

    Article  PubMed  Google Scholar 

  17. Mehrdad Mohammadpour ZH. Diagnostics in ocular imaging. Cham: Springer; 2021. p. 57.

    Book  Google Scholar 

  18. Lopes BT, Ramos IC, Salomão MQ, Guerra FP, Schallhorn SC, Schallhorn JM, et al. Enhanced tomographic assessment to detect corneal ectasia based on artificial intelligence. Am J Ophthalmol. 2018;195:223–32.

    Article  PubMed  Google Scholar 

  19. Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31(1):156–62.

    Article  PubMed  Google Scholar 

  20. Roberts CJ. Concepts and misconceptions in corneal biomechanics. J Cataract Refract Surg. 2014;40(6):862–9.

    Article  PubMed  Google Scholar 

  21. Piñero DP, Alcón N. In vivo characterization of corneal biomechanics. J Cataract Refract Surg. 2014;40(6):870–87.

    Article  PubMed  Google Scholar 

  22. Shah S, Laiquzzaman M, Cunliffe I, Mantry S. The use of the Reichert ocular response analyser to establish the relationship between ocular hysteresis, corneal resistance factor and central corneal thickness in normal eyes. Cont Lens Anterior Eye. 2006;29(5):257–62.

    Article  PubMed  Google Scholar 

  23. Schweitzer C, Roberts CJ, Mahmoud AM, Colin J, Maurice-Tison S, Kerautret J. Screening of forme fruste keratoconus with the ocular response analyzer. Invest Ophthalmol Vis Sci. 2010;51(5):2403–10.

    Article  PubMed  Google Scholar 

  24. Ayar O, Ozmen MC, Muftuoglu O, Akdemir MO, Koc M, Ozulken K. In-vivo corneal biomechanical analysis of unilateral keratoconus. Int J Ophthalmol. 2015;8(6):1141–5.

    PubMed  PubMed Central  Google Scholar 

  25. Zhang L, Danesh J, Tannan A, Phan V, Yu F, Hamilton DR. Second-generation corneal deformation signal waveform analysis in normal, forme fruste keratoconic, and manifest keratoconic corneas after statistical correction for potentially confounding factors. J Cataract Refract Surg. 2015;41(10):2196–204.

    Article  PubMed  Google Scholar 

  26. Mohammadpour M, Etesami I, Yavari Z, Naderan M, Abdollahinia F, Jabbarvand M. Ocular response analyzer parameters in healthy, keratoconus suspect and manifest keratoconus eyes. Oman J Ophthalmol. 2015;8(2):102–6.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Kirgiz A, Karaman Erdur S, Atalay K, Gurez C. The role of ocular response analyzer in differentiation of forme fruste keratoconus from corneal astigmatism. Eye Contact Lens. 2019;45(2):83–7.

    Article  PubMed  Google Scholar 

  28. Luz A, Lopes B, Hallahan KM, Valbon B, Ramos I, Faria-Correia F, et al. Enhanced combined tomography and biomechanics data for distinguishing forme fruste keratoconus. J Refract Surg. 2016;32(7):479–94.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Sedaghat MR, Momeni-Moghaddam H, Heravian J, Ansari A, Shayanfar H, Moshirfar M. Detection ability of corneal biomechanical parameters for early diagnosis of ectasia. Eye (Lond). 2023;37(8):1665–72.

    Article  PubMed  Google Scholar 

  30. Kirwan C, O’Malley D, O’Keefe M. Corneal hysteresis and corneal resistance factor in keratoectasia: findings using the reichert ocular response analyzer. Ophthalmologica. 2008;222(5):334–7.

    Article  PubMed  Google Scholar 

  31. Fraenkel D, Hamon L, Daas L, Flockerzi E, Suffo S, Eppig T, et al. Tomographically normal partner eye in very asymmetrical corneal ectasia: biomechanical analysis. J Cataract Refract Surg. 2021;47(3):366–72.

    Article  PubMed  Google Scholar 

  32. Atalay E, Özalp O, Erol MA, Bilgin M, Yildirim N. A combined biomechanical and tomographic model for identifying cases of subclinical keratoconus. Cornea. 2020;39(4):461–7.

    Article  PubMed  Google Scholar 

  33. Galletti JD, Ruiseñor Vazquez PR, Fuentes Bonthoux F, Pförtner T, Galletti JG. Multivariate analysis of the ocular response analyzer’s corneal deformation response curve for early keratoconus detection. J Ophthalmol. 2015;2015:496382.

    Article  PubMed  PubMed Central  Google Scholar 

  34. Steinberg J, Siebert M, Katz T, Frings A, Mehlan J, Druchkiv V, et al. Tomographic and biomechanical Scheimpflug imaging for keratoconus characterization: a validation of current indices. J Refract Surg. 2018;34(12):840–7.

    Article  PubMed  Google Scholar 

  35. Ambrósio R Jr, Lopes BT, Faria-Correia F, Salomao MQ, Buhren J, Roberts CJ, et al. Integration of Scheimpflug-based corneal tomography and biomechanical assessments for enhancing ectasia detection. J Refract Surg. 2017;33(7):434–43.

    Article  PubMed  Google Scholar 

  36. Maile H, Li JO, Gore D, Leucci M, Mulholland P, Hau S, et al. Machine learning algorithms to detect subclinical keratoconus: systematic review. JMIR Med Inform. 2021;9(12):e27363.

    Article  PubMed  PubMed Central  Google Scholar 

  37. Niazi S, Jimenez-Garcia M, Findl O, Gatzioufas Z, Doroodgar F, Shahriari MH, et al. Keratoconus diagnosis: from fundamentals to artificial intelligence: a systematic narrative review. Diagnostics (Basel). 2023;13(16):2715.

    Article  PubMed  PubMed Central  Google Scholar 

  38. Vandevenne MM, Favuzza E, Veta M, Lucenteforte E, Berendschot TT, Mencucci R, et al. Artificial intelligence for detecting keratoconus. Cochrane Database Syst Rev. 2023;11(11):CD014911.

    PubMed  Google Scholar 

  39. Rabinowitz YS, Rasheed K. KISA% index: a quantitative videokeratography algorithm embodying minimal topographic criteria for diagnosing keratoconus. J Cataract Refract Surg. 1999;25(10):1327–35.

    Article  PubMed  CAS  Google Scholar 

  40. McMahon TT, Szczotka-Flynn L, Barr JT, Anderson RJ, Slaughter ME, Lass JH, et al. A new method for grading the severity of keratoconus: the Keratoconus Severity Score (KSS). Cornea. 2006;25(7):794–800.

    Article  PubMed  Google Scholar 

  41. Amsler M. Classic keratocene and crude keratocene; unitary arguments. Ophthalmologica. 1946;111(2–3):96–101.

    Article  PubMed  CAS  Google Scholar 

  42. Valbon BF, Ambrósio R Jr, Fontes BM, Alves MR. Effects of age on corneal deformation by non-contact tonometry integrated with an ultra-high-speed (UHS) Scheimpflug camera. Arq Bras Oftalmol. 2013;76(4):229–32.

    Article  PubMed  Google Scholar 

  43. Esporcatte LPG, Salomão MQ, Lopes BT, Sena N, Ferreira E, Filho JBRF, et al. Biomechanics in keratoconus diagnosis. Curr Eye Res. 2023;48(2):130–6.

    Article  PubMed  Google Scholar 

  44. Vinciguerra R, Ambrósio R Jr, Elsheikh A, Roberts CJ, Lopes B, Morenghi E, et al. Detection of keratoconus with a new biomechanical index. J Refract Surg. 2016;32(12):803–10.

    Article  PubMed  Google Scholar 

  45. Roberts CJ, Mahmoud AM, Bons JP, Hossain A, Elsheikh A, Vinciguerra R, et al. Introduction of two novel stiffness parameters and interpretation of air puff-induced biomechanical deformation parameters with a dynamic Scheimpflug analyzer. J Refract Surg. 2017;33(4):266–73.

    Article  PubMed  Google Scholar 

  46. Eliasy A, Chen KJ, Vinciguerra R, Lopes BT, Abass A, Vinciguerra P, et al. Determination of corneal biomechanical behavior in-vivo for healthy eyes using CorVis ST tonometry: stress-strain index. Front Bioeng Biotechnol. 2019;7:105.

    Article  PubMed  PubMed Central  Google Scholar 

  47. Zhang H, Eliasy A, Lopes B, Abass A, Vinciguerra R, Vinciguerra P, et al. Stress-strain index map: a new way to represent corneal material stiffness. Front Bioeng Biotechnol. 2021;9:640434.

    Article  PubMed  PubMed Central  Google Scholar 

  48. Hwang ES, Perez-Straziota CE, Kim SW, Santhiago MR, Randleman JB. Distinguishing highly asymmetric keratoconus eyes using combined Scheimpflug and spectral-domain OCT analysis. Ophthalmology. 2018;125(12):1862–71.

    Article  PubMed  Google Scholar 

  49. Awad EA, Abou Samra WA, Torky MA, El-Kannishy AM. Objective and subjective diagnostic parameters in the fellow eye of unilateral keratoconus. BMC Ophthalmol. 2017;17(1):186.

    Article  PubMed  PubMed Central  Google Scholar 

  50. Shajari M, Jaffary I, Herrmann K, Grunwald C, Steinwender G, Mayer WJ, et al. Early tomographic changes in the eyes of patients with keratoconus. J Refract Surg. 2018;34(4):254–9.

    Article  PubMed  Google Scholar 

  51. Zhang D, Tian L, Zhang H, Zheng Y, Fu C, Zhai C, et al. Differences of corneal biomechanics among thin normal cornea, forme-fruste keratoconus, and cornea after SMILE. Front Bioeng Biotechnol. 2022;10:861924.

    Article  PubMed  PubMed Central  Google Scholar 

  52. Tian L, Qin X, Zhang H, Zhang D, Guo LL, Zhang HX, et al. A potential screening index of corneal biomechanics in healthy subjects, forme fruste keratoconus patients and clinical keratoconus patients. Front Bioeng Biotechnol. 2021;9:766605.

    Article  PubMed  PubMed Central  Google Scholar 

  53. Chen X, Cao H, Huo Y, Song J, Zou H, Li J, et al. Screening of sensitive in vivo characteristics for early keratoconus diagnosis: a multicenter study. Front Bioeng Biotechnol. 2023;11:1158299.

    Article  PubMed  PubMed Central  Google Scholar 

  54. Liu Y, Zhang Y, Chen Y. Application of a Scheimpflug-based biomechanical analyser and tomography in the early detection of subclinical keratoconus in Chinese patients. BMC Ophthalmol. 2021;21(1):339.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  55. Koc M, Aydemir E, Tekin K, Inanc M, Kosekahya P, Kiziltoprak H. Biomechanical analysis of subclinical keratoconus with normal topographic, topometric, and tomographic findings. J Refract Surg. 2019;35(4):247–52.

    Article  PubMed  Google Scholar 

  56. Xian Y, Zhao Y, Sun L, Zhang X, Ding L, Liu Z, et al. Comparison of bilateral differential characteristics of corneal biomechanics between keratoconus and normal eyes. Front Bioeng Biotechnol. 2023;11:1163223.

    Article  PubMed  PubMed Central  Google Scholar 

  57. Wang YM, Chan TCY, Yu M, Jhanji V. Comparison of corneal dynamic and tomographic analysis in normal, forme fruste keratoconic, and keratoconic eyes. J Refract Surg. 2017;33(9):632–8.

    Article  PubMed  Google Scholar 

  58. Miao YY, Ma XM, Qu ZX, Eliasy A, Wu B, Xu H, et al. Performance of updated stress-strain index in differentiating between normal, forme-fruste, subclinical and clinical keratoconic eyes. Am J Ophthalmol. 2024;258:196–207.

    Article  PubMed  CAS  Google Scholar 

  59. Zhang H, Zhang X, Hua L, Li L, Tian L, Zhang X, et al. An exploratory analysis of forme fruste keratoconus sensitivity diagnostic parameters. Int Ophthalmol. 2022;42(8):2473–81.

    Article  PubMed  Google Scholar 

  60. Huo Y, Chen X, Cao H, Li J, Hou J, Wang Y. Biomechanical properties analysis of forme fruste keratoconus and subclinical keratoconus. Graefes Arch Clin Exp Ophthalmol. 2023;261(5):1311–20.

    Article  PubMed  Google Scholar 

  61. Peris-Martínez C, Díez-Ajenjo MA, García-Domene MC, Pinazo-Durán MD, Luque-Cobija MJ, Del Buey-Sayas MÁ, et al. Evaluation of intraocular pressure and other biomechanical parameters to distinguish between subclinical keratoconus and healthy corneas. J Clin Med. 2021;10(9):1905.

    Article  PubMed  PubMed Central  Google Scholar 

  62. Steinberg J, Katz T, Lucke K, Frings A, Druchkiv V, Linke SJ. Screening for keratoconus with new dynamic biomechanical in vivo Scheimpflug analyses. Cornea. 2015;34(11):1404–12.

    Article  PubMed  Google Scholar 

  63. Castro-Luna G, Jiménez-Rodríguez D, Castaño-Fernández AB, Pérez-Rueda A. Diagnosis of subclinical keratoconus based on machine learning techniques. J Clin Med. 2021;10(18):4281.

    Article  PubMed  PubMed Central  Google Scholar 

  64. Ren S, Xu L, Fan Q, Gu Y, Yang K. Accuracy of new Corvis ST parameters for detecting subclinical and clinical keratoconus eyes in a Chinese population. Sci Rep. 2021;11(1):4962.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  65. Heidari Z, Hashemi H, Mohammadpour M, Amanzadeh K, Fotouhi A. Evaluation of corneal topographic, tomographic and biomechanical indices for detecting clinical and subclinical keratoconus: a comprehensive three-device study. Int J Ophthalmol. 2021;14(2):228–39.

    Article  PubMed  PubMed Central  Google Scholar 

  66. Chan TCY, Wang YM, Yu M, Jhanji V. Comparison of corneal tomography and a new combined tomographic biomechanical index in subclinical keratoconus. J Refract Surg. 2018;34(9):616–21.

    Article  PubMed  Google Scholar 

  67. Song Y, Feng Y, Qu M, Ma Q, Tian H, Li D, et al. Analysis of the diagnostic accuracy of Belin/Ambrósio enhanced ectasia and Corvis ST parameters for subclinical keratoconus. Int Ophthalmol. 2023;43(5):1465–75.

    Article  PubMed  Google Scholar 

  68. Mohammadpour M, Heidari Z, Hashemi H, Yaseri M, Fotouhi A. Comparison of artificial intelligence-based machine learning classifiers for early detection of keratoconus. Eur J Ophthalmol. 2022;32(3):1352–60.

    Article  PubMed  Google Scholar 

  69. Augustin VA, Son HS, Baur I, Zhao L, Auffarth GU, Khoramnia R. Detecting subclinical keratoconus by biomechanical analysis in tomographically regular keratoconus fellow eyes. Eur J Ophthalmol. 2021. https://doiorg.publicaciones.saludcastillayleon.es/10.1177/11206721211063740.

    Article  PubMed  Google Scholar 

  70. Herber R, Hasanli A, Lenk J, Vinciguerra R, Terai N, Pillunat LE, et al. Evaluation of corneal biomechanical indices in distinguishing between normal, very asymmetric, and bilateral keratoconic eyes. J Refract Surg. 2022;38(6):364–72.

    Article  PubMed  Google Scholar 

  71. Kataria P, Padmanabhan P, Gopalakrishnan A, Padmanaban V, Mahadik S, Ambrósio R Jr. Accuracy of scheimpflug-derived corneal biomechanical and tomographic indices for detecting subclinical and mild keratectasia in a South Asian population. J Cataract Refract Surg. 2019;45(3):328–36.

    Article  PubMed  Google Scholar 

  72. Wallace HB, Vellara HR, Gokul A, McGhee CNJ, Meyer JJ. Comparison of ectasia detection in early keratoconus using Scheimpflug-based corneal tomography and biomechanical assessments. Cornea. 2023;42(12):1528–35.

    Article  PubMed  Google Scholar 

  73. Salomão MQ, Hofling-Lima AL, Gomes Esporcatte LP, Lopes B, Vinciguerra R, Vinciguerra P, et al. The role of corneal biomechanics for the evaluation of ectasia patients. Int J Environ Res Public Health. 2020;17(6):2113.

    Article  PubMed  PubMed Central  Google Scholar 

  74. Ferreira-Mendes J, Lopes BT, Faria-Correia F, Salomão MQ, Rodrigues-Barros S, Ambrósio R Jr. Enhanced ectasia detection using corneal tomography and biomechanics. Am J Ophthalmol. 2019;197:7–16.

    Article  PubMed  Google Scholar 

  75. Koh S, Ambrósio R Jr, Inoue R, Maeda N, Miki A, Nishida K. Detection of subclinical corneal ectasia using corneal tomographic and biomechanical assessments in a Japanese population. J Refract Surg. 2019;35(6):383–90.

    Article  PubMed  Google Scholar 

  76. Augustin VA, Son HS, Kovalchuk B, Yildirim TM, Koppe MK, Auffarth GU, et al. Biomechanical analysis of tomographically regular keratoconus fellow eyes using Corvis ST. Klin Monbl Augenheilkd. 2023;240(8):944–51.

    PubMed  Google Scholar 

  77. Padmanabhan P, Lopes BT, Eliasy A, Abass A, Vinciguerra R, Vinciguerra P, et al. Evaluation of corneal biomechanical behavior in vivo for healthy and keratoconic eyes using the stress-strain index. J Cataract Refract Surg. 2022;48(10):1162–7.

    Article  PubMed  Google Scholar 

  78. Tian L, Zhang D, Guo L, Qin X, Zhang H, Zhang H, et al. Comparisons of corneal biomechanical and tomographic parameters among thin normal cornea, forme fruste keratoconus, and mild keratoconus. Eye Vis (Lond). 2021;8(1):44.

    Article  PubMed  Google Scholar 

  79. Guo LL, Tian L, Cao K, Li YX, Li N, Yang WQ, et al. Comparison of the morphological and biomechanical characteristics of keratoconus, forme fruste keratoconus, and normal corneas. Semin Ophthalmol. 2021;36(8):671–8.

    Article  PubMed  Google Scholar 

  80. Zhang H, Tian L, Guo L, Qin X, Zhang D, Li L, et al. Comprehensive evaluation of corneas from normal, forme fruste keratoconus and clinical keratoconus patients using morphological and biomechanical properties. Int Ophthalmol. 2021;41(4):1247–59.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  81. Wallace HB, Misra SL, Li SS, McKelvie J. Biomechanical changes in the cornea following cataract surgery: a prospective assessment with the Corneal Visualisation Scheimpflug Technology. Clin Exp Ophthalmol. 2019;47(4):461–8.

    Article  PubMed  Google Scholar 

  82. Roberts CJ, Dupps WJ Jr. Biomechanics of corneal ectasia and biomechanical treatments. J Cataract Refract Surg. 2014;40(6):991–8.

    Article  PubMed  PubMed Central  Google Scholar 

  83. Piñero DP, Nieto JC, Lopez-Miguel A. Characterization of corneal structure in keratoconus. J Cataract Refract Surg. 2012;38(12):2167–83.

    Article  PubMed  Google Scholar 

  84. Kenia VP, Kenia RV, Maru S, Pirdankar OH. Role of corneal epithelial mapping, Corvis biomechanical index, and artificial intelligence-based tomographic biomechanical index in diagnosing spectrum of keratoconus. Oman J Ophthalmol. 2023;16(2):276–80.

    Article  PubMed  PubMed Central  Google Scholar 

  85. Song P, Yang K, Li P, Liu Y, Liang D, Ren S, et al. Assessment of corneal pachymetry distribution and morphologic changes in subclinical keratoconus with normal biomechanics. Biomed Res Int. 2019;2019:1748579.

    Article  PubMed  PubMed Central  Google Scholar 

  86. Dong R, Liu Y, Zhang Y, Chen Y. The application of corneal biomechanical interocular asymmetry for the diagnosis of keratoconus and subclinical keratoconus. Front Bioeng Biotechnol. 2023;11:1266940.

    Article  PubMed  PubMed Central  Google Scholar 

  87. Vinciguerra R, Herber R, Wang Y, Zhang F, Zhou X, Bai J, et al. Corneal biomechanics differences between Chinese and Caucasian healthy subjects. Front Med. 2022;9:834663.

    Article  Google Scholar 

  88. Vinciguerra R, Ambrósio R, Wang Y, Zhang F, Zhou X, Bai J, et al. Detection of keratoconus with a new Corvis biomechanical index optimized for Chinese populations. Am J Ophthalmol. 2023;252:182–7.

    Article  PubMed  Google Scholar 

  89. Belin MW, Duncan JK. Keratoconus: the ABCD grading system. Klin Monbl Augenheilkd. 2016;233(6):701–7.

    Article  PubMed  CAS  Google Scholar 

  90. Yasuno Y, Madjarova VD, Makita S, Akiba M, Morosawa A, Chong C, et al. Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments. Opt Express. 2005;13(26):10652–64.

    Article  PubMed  Google Scholar 

  91. Izatt JA, Hee MR, Swanson EA, Lin CP, Huang D, Schuman JS, et al. Micrometer-scale resolution imaging of the anterior eye in vivo with optical coherence tomography. Arch Ophthalmol. 1994;112(12):1584–9.

    Article  PubMed  CAS  Google Scholar 

  92. Maeda N. Optical coherence tomography for corneal diseases. Eye Contact Lens. 2010;36(5):254–9.

    Article  PubMed  Google Scholar 

  93. Fan R, Chan TC, Prakash G, Jhanji V. Applications of corneal topography and tomography: a review. Clin Exp Ophthalmol. 2018;46(2):133–46.

    Article  PubMed  Google Scholar 

  94. Savini G, Schiano-Lomoriello D, Hoffer KJ. Repeatability of automatic measurements by a new anterior segment optical coherence tomographer combined with Placido topography and agreement with 2 Scheimpflug cameras. J Cataract Refract Surg. 2018;44(4):471–8.

    Article  PubMed  Google Scholar 

  95. Saad A, Debellemanière G, Zeboulon P, Rizk M, Rouger H, Mazharian A, et al. Discrimination between keratoconus, forme fruste keratoconus, and normal eyes using a novel OCT-based tomographer. J Cataract Refract Surg. 2023;49(11):1092–7.

    Article  PubMed  Google Scholar 

  96. Nakagawa T, Maeda N, Higashiura R, Hori Y, Inoue T, Nishida K. Corneal topographic analysis in patients with keratoconus using 3-dimensional anterior segment optical coherence tomography. J Cataract Refract Surg. 2011;37(10):1871–8.

    Article  PubMed  Google Scholar 

  97. Fujimoto H, Maeda N, Shintani A, Nakagawa T, Fuchihata M, Higashiura R, et al. Quantitative evaluation of the natural progression of keratoconus using three-dimensional optical coherence tomography. Invest Ophthalmol Vis Sci. 2016;57(9):OCT169-75.

    Article  PubMed  Google Scholar 

  98. Morishige N, Magome K, Ueno A, Matsui TA, Nishida T. Relations among corneal curvature, thickness, and volume in keratoconus as evaluated by anterior segment-optical coherence tomography. Invest Ophthalmol Vis Sci. 2019;60(12):3794–802.

    Article  PubMed  CAS  Google Scholar 

  99. Maeno S, Koh S, Inoue R, Oie Y, Maeda N, Jhanji V, et al. Fourier analysis on irregular corneal astigmatism using optical coherence tomography in various severity stages of keratoconus. Am J Ophthalmol. 2022;243:55–65.

    Article  PubMed  Google Scholar 

  100. Pavlatos E, Chen S, Chamberlain W, Huang D, Li Y. Detection of corneal ectasia using OCT maps of pachymetry and posterior surface mean curvature. J Refract Surg. 2022;38(8):502–10.

    Article  PubMed  PubMed Central  Google Scholar 

  101. Fukuda S, Beheregaray S, Hoshi S, Yamanari M, Lim Y, Hiraoka T, et al. Comparison of three-dimensional optical coherence tomography and combining a rotating Scheimpflug camera with a Placido topography system for forme fruste keratoconus diagnosis. Br J Ophthalmol. 2013;97(12):1554–9.

    Article  PubMed  Google Scholar 

  102. Itoi M, Kitazawa K, Yokota I, Wakimasu K, Cho Y, Nakamura Y, et al. Anterior and posterior ratio of corneal surface areas: a novel index for detecting early stage keratoconus. PLoS One. 2020;15(4):e0231074.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  103. Kitazawa K, Itoi M, Yokota I, Wakimasu K, Cho Y, Nakamura Y, et al. Involvement of anterior and posterior corneal surface area imbalance in the pathological change of keratoconus. Sci Rep. 2018;8(1):14993.

    Article  PubMed  PubMed Central  Google Scholar 

  104. Steinberg J, Casagrande MK, Frings A, Katz T, Druchkiv V, Richard G, et al. Screening for subclinical keratoconus using swept-source fourier domain anterior segment optical coherence tomography. Cornea. 2015;34(11):1413–9.

    Article  PubMed  Google Scholar 

  105. Shiga S, Kojima T, Nishida T, Nakamura T, Ichikawa K. Evaluation of CorvisST biomechanical parameters and anterior segment optical coherence tomography for diagnosing forme fruste keratoconus. Acta Ophthalmol. 2021;99(6):644–51.

    Article  PubMed  Google Scholar 

  106. Yang Y, Pavlatos E, Chamberlain W, Huang D, Li Y. Keratoconus detection using OCT corneal and epithelial thickness map parameters and patterns. J Cataract Refract Surg. 2021;47(6):759–66.

    Article  PubMed  PubMed Central  Google Scholar 

  107. Reinstein DZ, Archer TJ, Gobbe M. Corneal epithelial thickness profile in the diagnosis of keratoconus. J Refract Surg. 2009;25(7):604–10.

    Article  PubMed  Google Scholar 

  108. Reinstein DZ, Gobbe M, Archer TJ, Silverman RH, Coleman DJ. Epithelial, stromal, and total corneal thickness in keratoconus: three-dimensional display with artemis very-high frequency digital ultrasound. J Refract Surg. 2010;26(4):259–71.

    Article  PubMed  PubMed Central  Google Scholar 

  109. Li Y, Tan O, Brass R, Weiss JL, Huang D. Corneal epithelial thickness mapping by Fourier-domain optical coherence tomography in normal and keratoconic eyes. Ophthalmology. 2012;119(12):2425–33.

    Article  PubMed  Google Scholar 

  110. Feng Y, Reinstein DZ, Nitter T, Archer TJ, McAlinden C, Chen X, et al. Heidelberg Anterion Swept-Source OCT corneal epithelial thickness mapping: repeatability and agreement with Optovue Avanti. J Refract Surg. 2022;38(6):356–63.

    Article  PubMed  Google Scholar 

  111. Lu NJ, Hafezi F, Koppen C, Alió Del Barrio JL, Aslanides IM, Awwad ST, et al. New keratoconus staging system based on OCT. J Cataract Refract Surg. 2023;49(11):1098–105.

    Article  PubMed  Google Scholar 

  112. Wang Q, Lim L, Lim SWY, Htoon HM. Comparison of corneal epithelial and stromal thickness between keratoconic and normal eyes in an Asian population. Ophthalmic Res. 2019;62(3):134–40.

    Article  PubMed  Google Scholar 

  113. Heidari Z, Mohammadpour M, Hajizadeh F, Fotouhi A, Hashemi H. Corneal layer thickness in keratoconus using optical coherence tomography. Clin Exp Optom. 2024;107(1):32–9.

    Article  PubMed  Google Scholar 

  114. Li Y, Chamberlain W, Tan O, Brass R, Weiss JL, Huang D. Subclinical keratoconus detection by pattern analysis of corneal and epithelial thickness maps with optical coherence tomography. J Cataract Refract Surg. 2016;42(2):284–95.

    Article  PubMed  PubMed Central  Google Scholar 

  115. Temstet C, Sandali O, Bouheraoua N, Hamiche T, Galan A, El Sanharawi M, et al. Corneal epithelial thickness mapping using Fourier-domain optical coherence tomography for detection of form fruste keratoconus. J Cataract Refract Surg. 2015;41(4):812–20.

    Article  PubMed  Google Scholar 

  116. Koh S, Inoue R, Ambrósio R, Bernado L, Matsuo R, Maeno S, et al. Corneal layer thickness mapping with optical coherence tomography in different stages reveals pathologic processes of keratoconus. Cornea. 2024. https://doiorg.publicaciones.saludcastillayleon.es/10.1097/ICO.0000000000003693.

    Article  PubMed  Google Scholar 

  117. Feng Y, Reinstein DZ, Nitter T, Archer TJ, McAlinden C, Bertelsen G, et al. Epithelial thickness mapping in keratoconic corneas: repeatability and agreement between CSO MS-39, Heidelberg Anterion, and Optovue Avanti OCT devices. J Refract Surg. 2023;39(7):474–80.

    Article  PubMed  Google Scholar 

  118. Salomão MQ, Hofling-Lima AL, Gomes Esporcatte LP, Correa FF, Meneses EF, Li Y, et al. Corneal ectasia detection by epithelial pattern standard deviation from OCT. J Cataract Refract Surg. 2023;49(2):190–4.

    Article  PubMed  Google Scholar 

  119. Pircher N, Beer F, Holzer S, Gschließer A, Donner R, Pircher M, et al. Large field of view corneal epithelium and Bowman’s layer thickness maps in keratoconic and healthy eyes. Am J Ophthalmol. 2020;209:168–77.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

SH and SM were funded by the Medical Research Council (UK) (Grant No. MR/S037829/1); XRW was partly funded by Social Development Grant of Shaanxi Province (China) (Grant No. 2022SF-404) and fully funded by China Scholarship Council. YXW was funded by the China Scholarship Council. SM was funded in Cardiff by a Japan Partnership Award (BB/R021244/1) from the Biotechnology and Biological Research Council (UK) to Professor AJQ and Professor Kohji Nishida, Osaka University, Japan.

Author information

Authors and Affiliations

Authors

Contributions

XRW, SM, SK, SH and CM conceived the concept; XRW, SM, YXW, SK, AJQ, SH, SRM, CM wrote the manuscript; SM and SK collected data and performed data analysis; XRW, SM, SK, SH, and CM conducted revisions. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Colm McAlinden.

Ethics declarations

Ethics approval and consent to participate

The study was approval by the Institutional Review Board/Ethics Committee of Osaka University Hospital (registration number: 09297–20). Data was collected in accordance with the Declaration of Helsinki. Written informed consent was obtained from all participants or their parents for those under the age of 18 years.

Consent for publication

Not applicable.

Competing interests

Outside the submitted work, Dr McAlinden has received consultancy fees/honorarium/travel support (past 36 months) from: Atia Vision (Campbell, California, USA), Bausch and Lomb (Bridgewater, New Jersey, USA), Bayer (Leverkusen, Germany), British Society of Refractive Surgery (Oxford, UK), BVI (Liège, Belgium), Cutting Edge (Labége, France), Hoya (Frankfurt, Germany), Knowledge Gate Group (Copenhagen, Denmark), Johnson & Johnson Surgical Vision (Santa Ana, California, USA), Keio University (Tokyo, Japan), Medevise Consulting SAS (Strasbourg, France), Portuguese Society of Ophthalmology (Coimbra, Portugal), Rayner (Worthing, UK), ROHTO (Tokyo, Japan), Royal College of Ophthalmologists (London, UK), SightGlass vision (Menlo Park, California, USA), Scope (Crawley, UK), SpyGlass (Aliso Viejo, California, USA), Sun Yat-sen University (Guangzhou, China), Thea pharmaceuticals (Clemont-Ferrand, France), Vold Vision (Arkansas, USA). Dr McAlinden developed the Quality of Vision (QoV) questionnaire and the Orthokeratology and Contact Lens Quality of Life Questionnaire (OCL-QoL), and has a financial interest in these tools. He also consults on topics including Rasch analysis, questionnaires, statistical analyses, and clinical/surgical ophthalmology topics. Dr McAlinden has been co-awarded research grants relating to diabetic eye disease from the Welsh Government and the Royal College of Ophthalmologists (unpaid roles) and Catalytic funding from Health and Care Research Wales to develop a Centre for Vision Services Research. Dr McAlinden is a council member and treasurer of the British Society for Refractive Surgery (unpaid role) and a PROM advisor to the Royal College of Ophthalmologists (unpaid role). Dr McAlinden has undertaken paid peer reviews for Research Square (Durham, North Carolina, USA). Dr McAlinden is an editorial board member for Graefe's Archive for Clinical and Experimental Ophthalmology, Eye and Vision, Archives of Medical Science, Journal of Clinical Medicine, Journal of Ophthalmology, and Journal of Clinical and Experimental Ophthalmology. Dr McAlinden is an Associate Editor for Frontiers in Medicine—Ophthalmology.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wang, X., Maeno, S., Wang, Y. et al. Early diagnosis of keratoconus using corneal biomechanics and OCT derived technologies. Eye and Vis 12, 18 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s40662-025-00435-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s40662-025-00435-3

Keywords