Table 2 Monocular and binocular postoperative values of visual acuity. All values are expressed as logMAR and Snellen

From: Influence of preoperative variables on the 3-month functional outcomes of the Vivity extended depth-of-focus intraocular lens: a prospective case series

Parameters

Mean ± SD LogMAR Snellen

Range LogMAR Snellen

95% CI

Monocular UCDVA

 − 0.02 ± 0.11 20/19

 − 0.10, + 0.63 20/32, 20/16

[− 0.04, 0.00]

Binocular UCDVA

 − 0.08 ± 0.05 20/17

 − 0.20, + 0.00 20/27, 20/13

[− 0.09, − 0.07]

Monocular DCVA

 − 0.06 ± 0.07 20/18

 − 0.10, + 0.20 20/32, 20/16

[− 0.07, − 0.04]

Binocular DCVA

 − 0.07 ± 0.06 20/17

 − 0.20, + 0.00 20/25, 20/13

[− 0.08, − 0.06]

Monocular DCIVA

0.06 ± 0.13 20/23

 − 0.20, + 0.50 20/63, 20/13

[0.04, 0.09]

Binocular DCIVA

0.01 ± 0.08 20/20

 − 0.20, + 0.20 20/32, 20/13

[− 0.01, 0.03]

Monocular DCNVA

0.30 ± 0.16 20/40

0.00, + 1.00 20/200, 20/20

[0.27, 0.34]

Binocular DCNVA

0.19 ± 0.13 20/31

 − 0.10, + 0.60 20/80, 20/16

[0.17, 0.22]

  1. SD = standard deviation; CI = confidence interval; UCDVA = uncorrected distance visual acuity; DCVA = distance-corrected visual acuity; DCIVA = distance-corrected intermediate visual acuity; DCNVA = distance-corrected near visual acuity