Skip to main content

Table 5 Results of agreement analysis between the Colombo IOL and IOLMaster 700 measurements

From: Repeatability and reproducibility of a new spectral-domain optical coherence tomography biometer and agreement with swept-source optical coherence tomography based biometer

Parameter

Mean difference ± SD

P value

95% LoA

AL (mm)

 − 0.03 ± 0.03

0.000

 − 0.08 to 0.03

CCT (μm)

 − 13.34 ± 4.21

0.000

 − 21.58 to − 5.09

AQD (mm)

0.08 ± 0.04

0.000

0.01 to 0.15

ACD (mm)

0.07 ± 0.04

0.000

 − 0.01 to 0.14

LT (mm)

0.02 ± 0.04

0.000

 − 0.05 to 0.10

Kf (D)

0.22 ± 0.19

0.000

 − 0.14 to 0.59

Ks (D)

0.04 ± 0.18

0.012

 − 0.31 to 0.40

Km (D)

0.13 ± 0.13

0.000

 − 0.13 to 0.40

AST (D)

 − 0.18 ± 0.25

0.000

 − 0.68 to 0.32

J0 (D)

0.12 ± 0.11

0.000

 − 0.09 to 0.34

J45 (D)

0.09 ± 0.08

0.000

 − 0.07 to 0.25

WTW (mm)

0.68 ± 0.40

0.000

 − 0.11 to 1.47

PD (mm)

0.67 ± 0.84

0.000

 − 0.97 to 2.31

  1. AL = axial length; CCT = central corneal thickness; AQD = aqueous humor depth; ACD = anterior chamber depth; LT = lens thickness; Kf = flattest meridian keratometry; Ks = steepest meridian keratometry; Km = mean keratometry; AST = corneal astigmatism; J0 = astigmatism parameters J0; J45 = astigmatism parameters J45; WTW = white to white; PD = pupil diameter; SD = standard deviation; LoA = limits of agreement