Skip to main content

Table 4 Results of reproducibility analysis of the Colombo IOL measurements in normal subjects

From: Repeatability and reproducibility of a new spectral-domain optical coherence tomography biometer and agreement with swept-source optical coherence tomography based biometer

Parameter

Mean ± SD

Sw

TRT

CoV (%)

ICC (95% CI)

AL (mm)

25.51 ± 1.14

0.00

0.01

0.02

1.000 (1.000–1.000)

CCT (μm)

531.41 ± 29.52

0.90

2.49

0.17

0.999 (0.998–0.999)

AQD (mm)

3.14 ± 0.31

0.02

0.06

0.70

0.995 (0.992–0.997)

ACD (mm)

3.67 ± 0.31

0.02

0.06

0.60

0.995 (0.992–0.997)

LT (mm)

3.81 ± 0.34

0.03

0.08

0.72

0.994 (0.989–0.996)

Kf (D)

43.12 ± 1.37

0.09

0.24

0.20

0.996 (0.994–0.998)

Ks (D)

44.14 ± 1.51

0.08

0.21

0.18

0.997 (0.996–0.998)

Km (D)

43.63 ± 1.41

0.07

0.19

0.16

0.998 (0.996–0.999)

AST (D)

1.02 ± 0.6

0.09

0.24

8.43

0.980 (0.967–0.987)

J0 (D)

 − 0.38 ± 0.37

0.05

0.13

 − 11.77

0.985 (0.977–0.991)

J45 (D)

0.14 ± 0.2

0.04

0.10

26.70

0.964 (0.942–0.978)

WTW (mm)

12.63 ± 0.54

0.09

0.24

0.70

0.973 (0.957–0.984)

PD (mm)

5.93 ± 1.05

0.34

0.95

5.77

0.900 (0.842–0.937)

  1. AL = axial length; CCT = central corneal thickness; AQD = aqueous humor depth; ACD = anterior chamber depth; LT = lens thickness; Kf = flattest meridian keratometry; Ks: steepest meridian keratometry; Km = mean keratometry; AST = corneal astigmatism; J0 = astigmatism parameters J0; J45 = astigmatism parameters J45; WTW = white to white; PD = pupil diameter; SD = standard deviation; Sw = within-subject standard deviation; TRT = test–retest repeatability coefficient (2.77 × Sw); CoV = coefficient of variation; ICC = intraclass correlation coefficient; CI = confidence interval